Using Laser Doppler Imaging and Monitoring to Analyze Spinal Cord Microcirculation in Rat
Published on: May 30, 2018
A technique based on laser Doppler flowmetry and photoplethysmography for simultaneously monitoring blood flow at
J Hagblad1, L-G Lindberg, A Kaisdotter Andersson
1School of innovation, Design and Engineering, Mälardalen University, Box 883, 721 23, Västerås, Sweden. jimmie.hagblad@mdh.se
Abstract:
The aim of this study was to validate a non-invasive optical probe for simultaneous blood flow measurement at different vascular depths combining three photoplethysmography (PPG) channels and laser Doppler flowmeter (LDF). Wavelengths of the PPG were near-infrared 810 nm with source-to-detector separation of 10 and 25 mm, and green 560 nm with source-to-detector separation of 4 mm. The probe is intended for clinical studies of pressure ulcer aetiology. The probe was placed over the trapezius muscle, and depths from the skin to the trapezius muscle were measured using ultrasound and varied between 3.8 and 23 mm in the 11 subjects included. A provocation procedure inducing a local enhancement of blood flow in the trapezius muscle was used. Blood flows at rest and post-exercise were compared. It can be concluded that this probe is useful as a tool for discriminating between blood flows at different vascular tissue depths. The vascular depths reached for the different channels in this study were at least 23 mm for the near-infrared PPG channel (source-to-detector separation 25 mm), 10-15 mm for the near-infrared PPG channel (separation 10 mm), and shallower than 4 mm for both the green PPG channel (separation 4 mm) and LDF.
Related Concept Videos
Blood Flow
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...