Related Concept Videos

Accuracy and Precision01:52

Accuracy and Precision

Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.  Highly accurate measurements...
Instrument Calibration01:12

Instrument Calibration

Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Uncertainty in Measurement: Accuracy and Precision03:37

Uncertainty in Measurement: Accuracy and Precision

Scientists typically make repeated measurements of a quantity to ensure the quality of their findings and to evaluate both the precision and the accuracy of their results. Measurements are said to be precise if they yield very similar results when repeated in the same manner. A measurement is considered accurate if it yields a result that is very close to the true or the accepted value. Precise values agree with each other; accurate values agree with a true value.
Distance Corrections01:15

Distance Corrections

To achieve precise distance measurements, especially in surveying and construction, certain corrections must be applied to account for potential sources of error like the standardization errors, temperature variations, and slope adjustments.Standardization error emerges when measurement equipment undergoes changes, such as wear, repairs, or weather impacts. To address this, surveyors compare the equipment’s readings to a standard. This process identifies any deviation that might lead to...
Glassware Calibration01:11

Glassware Calibration

Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Uncertainty in Measurement: Reading Instruments02:46

Uncertainty in Measurement: Reading Instruments

Counting is the type of measurement that is free from uncertainty, provided the number of objects being counted does not change during the process. Such measurements result in exact numbers. By counting the eggs in a carton, for instance, one can determine exactly how many eggs are there in the carton. Similarly, the numbers of defined quantities are also exact. For example, 1 foot is exactly 12 inches, 1 inch is exactly 2.54 centimeters, and 1 gram is exactly 0.001 kilograms. Quantities...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Asphyxial depolarisation in the spinal cord.

The American journal of physiology·2010
Same author

Multiply diffracted light in the czerny-turner spectrometer.

Applied optics·2010
Same author

Comments on "principles of infrared spectrophotometry.".

Science (New York, N.Y.)·1951
Same author

CHOLINESTERASES.

Science (New York, N.Y.)·1944
Same author

AN ELECTRONIC RELAY WITH IMPROVED CHARACTERISTICS.

Science (New York, N.Y.)·1941
Same journal

Multifunctional reconfigurable terahertz metasurface based on vanadium dioxide phase transition: achieving broadband absorption and efficient polarization conversion.

Applied optics·2026
Same journal

High-Q-factor electromagnetically induced transparency utilizing quasi-bound states in the continuum in an all-dielectric terahertz metasurface.

Applied optics·2026
Same journal

Automated stitching interferometry for high-precision metrology of X-ray mirrors.

Applied optics·2026
Same journal

Experimental demonstration of an approach to designing a metal-dielectric DBR resonant cavity structure.

Applied optics·2026
Same journal

High-precision wavefront reconstruction from a single-shot interferogram using a physics-driven hybrid feature calibration network.

Applied optics·2026
Same journal

Ultra-high-Q Fano resonance based on coupled topological corner states in Kagome photonic crystals.

Applied optics·2026
See all related articles

Related Experiment Video

Updated: Jun 16, 2026

Indoor Experimental Assessment of the Efficiency and Irradiance Spot of the Achromatic Doublet on Glass (ADG) Fresnel Lens for Concentrating Photovoltaics
09:00

Indoor Experimental Assessment of the Efficiency and Irradiance Spot of the Achromatic Doublet on Glass (ADG) Fresnel Lens for Concentrating Photovoltaics

Published on: October 27, 2017

Technique for measuring photometric accuracy.

R C Hawes

    Applied Optics
    |January 30, 2010
    PubMed
    Summary
    This summary is machine-generated.

    A new photometric linearity test offers high accuracy and practicality by combining Bouguer

    More Related Videos

    A Stable Phantom Material for Optical and Acoustic Imaging
    04:54

    A Stable Phantom Material for Optical and Acoustic Imaging

    Published on: June 16, 2023

    Measuring the Behavioral Effects of Intraocular Scatter
    05:10

    Measuring the Behavioral Effects of Intraocular Scatter

    Published on: February 18, 2021

    Related Experiment Videos

    Last Updated: Jun 16, 2026

    Indoor Experimental Assessment of the Efficiency and Irradiance Spot of the Achromatic Doublet on Glass (ADG) Fresnel Lens for Concentrating Photovoltaics
    09:00

    Indoor Experimental Assessment of the Efficiency and Irradiance Spot of the Achromatic Doublet on Glass (ADG) Fresnel Lens for Concentrating Photovoltaics

    Published on: October 27, 2017

    A Stable Phantom Material for Optical and Acoustic Imaging
    04:54

    A Stable Phantom Material for Optical and Acoustic Imaging

    Published on: June 16, 2023

    Measuring the Behavioral Effects of Intraocular Scatter
    05:10

    Measuring the Behavioral Effects of Intraocular Scatter

    Published on: February 18, 2021

    Area of Science:

    • Analytical Chemistry
    • Spectroscopy

    Background:

    • Photometric linearity is crucial for accurate spectrophotometer measurements.
    • Existing methods for assessing photometric linearity often lack accuracy or practicality.
    • There is a need for an improved, reliable method for linearity testing.

    Purpose of the Study:

    • To describe a convenient and accurate photometric linearity test.
    • To address the limitations of commonly used linearity testing methods.

    Main Methods:

    • The described method utilizes direct application of physical principles.
    • It combines Bouguer's law with the superposition of optical fields.
    • A commercial spectrophotometer with a calibrated potentiometer was used.

    Main Results:

    • The novel method demonstrates high accuracy and practicality.
    • Peak errors in absorbance were found to be 0.0011 or less.
    • Observed errors were primarily attributed to phototube limitations.

    Conclusions:

    • The combined approach of Bouguer's law and optical superposition provides a superior linearity test.
    • This method overcomes deficiencies present in alternative techniques.
    • The test is suitable for routine use in spectrophotometry.