Subwavenumber charge-coupled device spectrometer calibration using molecular iodine laser-induced fluorescence
Joseph G Lambert1, Carlos Hernandez-Diaz, J Charles Williamson
1Department of Physics, Drexel University, Philadelphia, Pennsylvania 19104, USA.
Abstract:
Spectrometers configured with charge-coupled devices (CCD) or other array-based detectors require calibration to convert from the pixel coordinate to a spectral coordinate. A CCD calibration method well suited for Raman spectroscopy has been developed based on the 514.5 nm Ar(+) laser-induced fluorescence (LIF) spectrum of room-temperature molecular iodine vapor. Over 360 primary and secondary I(2) LIF calibration lines spanning 510-645 nm were identified as calibrant peaks using an instrumental resolution of 1 cm(-1). Two instrument calibration functions were evaluated with these peaks: a second-order polynomial and a function derived from simple optomechanical considerations. The latter function provided better fitting characteristics. Calibration using I(2) LIF was tested with measurements of both laser light scattering and Raman spectra. The I(2) LIF reference spectra and the signal spectra were recorded simultaneously, with no cross talk, by separating the two signals spatially along the vertical axis of the CCD imager. In this way, every CCD image could be independently calibrated. An accuracy and a precision of +/-0.05 cm(-1) were achieved with this calibration technique.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
UV–Vis Spectroscopy: Molecular Electronic Transitions
IR Spectroscopy: Molecular Vibration Overview
Stretching vibrations are vibrational motions that occur along the bond line, changing the bond length or distance between two bonded atoms. They are further distinguished as symmetric or asymmetric. In symmetric stretching, the...