Related Experiment Video
Updated: Jun 16, 2026

Measurement of Dynamic Scapular Kinematics Using an Acromion Marker Cluster to Minimize Skin Movement Artifact
Published on: February 10, 2015
Subacromial morphometric assessment of the clavicle hook plate
Amr W ElMaraghy1, Moira W Devereaux, Kajeandra Ravichandiran
1St. Joseph's Health Centre, Department of Orthopaedic Surgery, Toronto, Ontario, Canada.
Background:
Clavicle hook plates are an effective plate fixation alternative for distal clavicle fractures and severe acromioclavicular joint dislocations. However, post-operative complications associated with the subacromial portion of the hook include acromial osteolysis and subacromial impingement. We examine and quantify the three-dimensional position of the subacromial portion of the hook plate relative to surrounding acromial and subacromial structures in a series of cadaveric shoulders to determine if hook positioning predisposes the shoulder to these noted post-operative complications.
Materials And Methods:
Fifteen cadaveric shoulders (seven males, eight females) were implanted with 15- or 18-mm hook plates. Dimensions of the acromion and hook plate were digitised and reconstructed into a three-dimensional model to measure acromion dimensions and distances of the subacromial hook relative to surrounding acromial and subacromial structures.
Results:
Inter-specimen dimensions of the acromion were highly variable. Mean acromion width and thickness were greater in males than in females (p=0.01). The posterior orientation of the subacromial hook varied widely (mean posterior implantation angle=32.5+/-20 degrees, range 0-67 degrees). The hook pierced the subacromial bursa in 13/15 specimens, made contact with the belly of the supraspinatus muscle in 9/15 specimens, and had focal contact at the hook tip with the undersurface of the acromion in 9/15 specimens.
Conclusions:
The wide range of acromial dimensions leads to a high degree of variability in the positioning of the subacromial hook. The observed frequency of hook contact with surrounding subacromial structures in a static shoulder confirms that the position of the hook portion of the implant can predispose anatomic structures to the post-operative complications of subacromial impingement and bony erosion.
Related Concept Videos
Muscles of the Shoulder
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...
Bones of the Upper Limb: Humerus
Muscles that Move the Arm
The pectoralis major has two origins. Its clavicular head originates on the medial half of the clavicle. In contrast, the sternocostal head originates on the costal cartilages of ribs 1-6, the sternum, and the aponeurosis of the external oblique of the...
Changes in the Appendicular Skeleton with Age
Initially, the limb buds consist of a core of mesenchyme covered by a layer of ectoderm. The ectoderm at the end of the limb bud thickens to form a narrow crest called the apical ectodermal ridge. This ridge stimulates the underlying...
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a short...