Hemodynamic impact of anastomosis size and angle in side-to-end arteriovenous fistulae: a computer analysis

Koen Van Canneyt1, Thierry Pourchez, Sunny Eloot

  • 1Institute of Biomedical Technology (IBiTech), Ghent University, Ghent, Belgium. Koen.VanCanneyt@UGent.be

Insights

This study reveals that arteriovenous fistula (AVF) anastomosis size and angle significantly impact hemodynamics. Larger openings and wider angles improve flow, while sharp angles can reverse arterial flow, crucial for hemodialysis access.

Area of Science:

  • Biomedical Engineering
  • Cardiovascular Physiology
  • Medical Device Design

Background:

  • Arteriovenous fistula (AVF) is the preferred vascular access for hemodialysis.
  • AVF complications often overlook the anastomosis (connection segment).
  • Hemodynamic factors influencing AVF function require further investigation.

Purpose of the Study:

  • To mathematically investigate the hemodynamic impact of anastomosis size and angle on pressure drop and flow distribution in AVFs.
  • To analyze how variations in anastomosis geometry affect blood flow dynamics within the access.

Main Methods:

  • A computational model of a side-to-end AVF was developed.
  • Substudies focused on varying anastomosis cross-sectional area and angle.
  • Pressure drop and flow distribution were calculated across a range of arterial inflow rates (600-1200 mL/min).

Main Results:

  • Increased anastomosis area and angles wider than 43 degrees reduced pressure drop.
  • Larger anastomosis areas shifted flow predominantly to the proximal vein.
  • Anastomosis angles exceeding 58 degrees led to insufficient flow and potential distal arterial flow reversal.

Conclusions:

  • Anastomosis size and angle are critical, underestimated factors influencing AVF hemodynamics.
  • These geometric parameters significantly affect pressure drop and flow distribution in AVFs.
  • Optimizing anastomosis geometry could improve AVF function and reduce complications.
Abstract