Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
Auditing the technology used to measure blood pressure
J N Amoore1, M Guehenec, R Scordecchia
1Department of Medical Physics, Royal Infirmary of Edinburgh, 51 Little France Crescent, Edinburgh, UK. john.amoore@luht.scot.nhs.uk
Abstract:
Patient management requires accurate blood pressure measurement, but there is concern about the accuracy of the equipment. Makes and types of sphygmomanometers and sizes of their associated cuffs used in hospital wards and outpatient departments were recorded and calibration checked. Half the sphygmomanometers were oscillometric, 32% aneroid and 18% mercury. There were faults in 30% of the aneroid devices and 32% of the mercury devices, mostly of cuffs, hoses and connectors. Of the aneroid devices 18% had calibration errors (mostly of only 4 mmHg at pressures above 200 mmHg). Nearly 90% of the aneroid devices had only one cuff size available, suggesting that miss-cuffing might be a serious cause of inaccurate blood pressure measurements. Many oscillometric devices had no evidence of compliance with validation standards. Sphygmomanometers and their cuffs and hoses should be regularly checked. Clinical staff can carry out many of the checks.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...