Auditing the technology used to measure blood pressure
J N Amoore1, M Guehenec, R Scordecchia
1Department of Medical Physics, Royal Infirmary of Edinburgh, 51 Little France Crescent, Edinburgh, UK. john.amoore@luht.scot.nhs.uk
Accurate blood pressure measurement is vital, but equipment accuracy is a concern. Many aneroid and mercury sphygmomanometers showed faults, and most aneroid devices lacked varied cuff sizes, potentially leading to inaccurate readings.
Area of Science:
- Clinical Measurement
- Medical Device Quality
- Cardiovascular Health
Background:
- Accurate blood pressure measurement is crucial for effective patient management.
- Concerns exist regarding the reliability and accuracy of commonly used sphygmomanometers.
- Variations in device types and cuff sizes may impact measurement precision.
Purpose of the Study:
- To assess the accuracy and identify common faults in sphygmomanometers used in clinical settings.
- To evaluate the prevalence of calibration errors and cuff size availability in aneroid devices.
- To examine the compliance of oscillometric devices with validation standards.
Main Methods:
- A survey of sphygmomanometer makes, types, and cuff sizes in hospital wards and outpatient departments.
- Calibration checks were performed on aneroid and mercury sphygmomanometers.
- Assessment of faults in cuffs, hoses, and connectors.
- Evaluation of oscillometric device compliance with validation standards.
Main Results:
- 30% of aneroid and 32% of mercury sphygmomanometers exhibited faults, primarily in cuffs, hoses, and connectors.
- 18% of aneroid devices showed calibration errors, particularly at higher pressures (>200 mmHg).
- Nearly 90% of aneroid devices had only one cuff size, increasing the risk of incorrect cuffing and inaccurate readings.
- Many oscillometric devices lacked documented compliance with validation standards.
Conclusions:
- Regular checking and maintenance of sphygmomanometers, cuffs, and hoses are essential for accurate blood pressure measurement.
- Clinical staff can perform many diagnostic checks to ensure device accuracy.
- Addressing issues like limited cuff size availability and device calibration is critical to improve blood pressure measurement reliability.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...