Related Experiment Video
Updated: Jun 16, 2026

Accuracy in Dental Medicine, A New Way to Measure Trueness and Precision
Published on: April 29, 2014
Accuracy of linear measurements from cone-beam computed tomography-derived surface models of different voxel sizes
Janalt Damstra1, Zacharias Fourie, James J R Huddleston Slater
1Department of Orthodontics, University Medical Center Groningen, University of Groningen, PO Box 30.001, Groningen, 9700 RB, The Netherlands. j.damstra@dmo.umcg.nl
Introduction:
The aims of this study were to determine the linear accuracy of 3-dimensional surface models derived from a commercially available cone-beam computed tomography (CBCT) dental imaging system and volumetric rendering software and to investigate the influence of voxel resolution on the linear accuracy of CBCT surface models.
Methods:
Glass sphere markers were fixed on 10 dry mandibles. The mandibles were scanned with 0.40 and 0.25 voxel size resolutions in 3 sessions. Anatomic truth was established with 6 direct digital caliper measurements. The surface models were rendered by a volumetric rendering program, and the CBCT measurements were established as the mean of the 3 measurements.
Results:
The intraclass correlation coefficients between the physical measurements and the measurements of the CBCT images of 0.40 and 0.25 voxels were all more than 0.99. All CBCT measurements were accurate. There was no difference between the accuracy of the measurements between the 0.40 and 0.25 voxel size groups. The smallest detectable differences of the CBCT measurements were minimal, confirming the accuracy of the CBCT measurement procedure.
Conclusions:
The measurements on 3-dimensional surface models of 0.25 and 0.40 voxel size data sets made with the 3D eXam CBCT scanner (KaVo Dental GmbH, Bismarckring, Germany) and SimPlant Ortho Pro software (version 2.00, Materialise Dental, Leuven, Belgium) are accurate compared with direct caliper measurements. An increased voxel resolution did not result in greater accuracy of the surface model measurments.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Linear Approximations
Imaging Studies III: Computed Tomography
Accuracy, limits, and approximation
Accuracy is defined as the closeness of the measured value to the true or actual value. In engineering mechanics, repeated measurements are taken during theoretical or experimental analyses to ensure that the result is precise and accurate.
The accuracy of any solution is based on the...