Varioilluminator (subtractive double monochromator with variable bandpass).
Applied Optics
|February 4, 2010
Summary
A novel varioilluminator, a type of double monochromator, offers adjustable wavelength bandpasses. This instrument allows independent selection of long and short wavelength cutoffs for versatile spectral filtering.
Area of Science:
- Optics and Photonics
- Spectroscopy Instrumentation
Background:
- Traditional monochromators often have fixed bandpass limitations.
- Achieving variable and precisely controlled spectral filtering can be challenging.
Purpose of the Study:
- To introduce a novel double monochromator design, the varioilluminator.
- To demonstrate a method for achieving independently selectable wavelength cutoffs.
- To enable a wide range of tunable bandpasses for spectral applications.
Main Methods:
- The varioilluminator utilizes subtractive dispersion in its second stage.
- A wide, independently adjustable central slit allows for bandpass control.
- The instrument can incorporate either quartz or glass prisms.
Main Results:
- The wavelength bandpass is primarily determined by the central slit width.
- Independent adjustment of the central slit jaws allows precise selection of long and short wavelength cutoffs.
- Bandpasses ranging from approximately 15 Angstroms to several hundred Angstroms are achievable.
Conclusions:
- The varioilluminator provides a flexible and tunable spectral filtering solution.
- Independent control over wavelength cutoffs enhances its utility in various spectroscopic applications.
- This design offers a significant advancement in variable bandpass monochromator technology.
Related Concept Videos
UV–Vis Spectrometers
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell. Samples for...
Phase Contrast and Differential Interference Contrast Microscopy
Phase-Contrast Microscopes
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
IR Spectrometers
There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...
Atomic Absorption Spectroscopy: Instrumentation
An atomic absorption spectrophotometer (AAS) comprises several components: a radiation source, an atomizer, a monochromator, and a detector. The radiation source can be a hollow-cathode lamp (HCL) or an electrodeless-discharge lamp (EDL), both of which provide a narrow emission line of the required wavelength. However, some instruments use continuum sources and high-resolution monochromators to achieve a narrow range of radiation.
The atomizer used in AAS can be either a flame atomizer or an...
The atomizer used in AAS can be either a flame atomizer or an...