Related Experiment Video
Updated: Jun 16, 2026

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Diagnostic utility of bipolar precordial leads during ajmaline testing for suspected Brugada syndrome
Velislav N Batchvarov1, Malini Govindan, Peter Macfarlane
1St. George's University of London, London, UK. vbatchva@sgul.ac.uk
Background:
Leads V(1) and V(2) recorded from the standard position (fourth intercostal space) have insufficient sensitivity to detect the diagnostic type 1 Brugada ECG pattern.
Objective:
The purpose of this study was to compare the sensitivity of bipolar leads with a positive pole at V(2) and a negative pole at V(4) or V(5) with that of the standard unipolar lead V(2) for detection of the type 1 Brugada pattern.
Methods:
We analyzed digital 15-lead ECGs (12 standard leads plus leads V(1) to V(3) recorded from the third intercostal space [V(1h) to V(3h)]) acquired during diagnostic ajmaline testing in 128 patients (80 men, age 37 +/- 15 years) with suspected Brugada syndrome and standard 12-lead ECGs recorded in 229 healthy subjects (111 men, age 33 +/- 4 years). Bipolar leads between V(2) (positive pole) and V(4) or V(5) (leads V(2-4), V(2-5)) were derived by subtracting leads V(4) and V(5) from V(2). All ECGs were examined for the presence of type 1 Brugada pattern.
Results:
During 21 (16.4%) positive ajmaline tests, type 1 pattern was observed in lead V(2h) during 20 tests (95.2%) and in V(2) during 10 tests (47.6%). Type 1 pattern appeared in lead V(2-4) or V(2-5) in all tests when it was present in V(2) and in seven tests during which it was observed in lead V(2h) but not V(2) (17 tests [81%]). Type 1-like pattern was observed in lead V(2-4) or V(2-5) during two nonpositive tests (1.9%) and in one healthy subject (0.4%).
Conclusion:
Bipolar leads V(2-4) and V(2-5) are more sensitive than lead V(2) for detection of the type 1 Brugada pattern.
More Related Videos
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
09:20Microelectrode Array Recording of Sinoatrial Node Firing Rate to Identify Intrinsic Cardiac Pacemaking Defects in Mice
Published on: July 5, 2021
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Dysrhythmias V: Evaluating Dysrhythmias
Mitral Stenosis II: Clinical features and Diagnostic Tests
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Mitral Regurgitation II: Clinical Features and Diagnostic Tests