Contrast detail phantom comparison on a commercially available unit. Digital breast tomosynthesis (DBT) versus
Marco Bertolini1, Andrea Nitrosi, Giovanni Borasi
1Servizio di Fisica Medica Sanitaria, Azienda Ospedaliera S. Maria Nuova, Reggio Emilia, Italy.
Abstract:
The performance of a commercial digital mammographic system working in 2D planar versus tomosynthesis mode was evaluated in terms of the image signal difference to noise ratio (SDNR). A contrast detail phantom was obtained embedding 1 cm Plexiglas, including 49 holes of different diameter and depth, between two layers containing a breast-simulating material. The phantom was exposed with the details plane perpendicular to the X-ray beam using the manufacturer's standard clinical breast acquisition parameters. SDNR in the digital breast tomosynthesis (DBT) images was higher than that of the full-field digital mammography (FFDM) for 38 out of 49 details in complex background conditions. These differences (p < 0.05) are statistically significant for 19 details out of 38. The relative SDNR results for DBT and FFDM images showed a dependence on the diameter of the details considered. This paper proposes an initial framework for a global image quality evaluation for commercial systems that can operate with different image acquisition modality using the same detector.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...