Published on: November 23, 2019
High-resolution cartilage imaging of the knee at 3T: basic evaluation of modern isotropic 3D MR-sequences
Klaus M Friedrich1, Gert Reiter, Bernd Kaiser
1MR Centre of Excellence, Division of Neuroradiology and Musculoskeletal Radiology, Department of Radiology, Medical University of Vienna, Waehringer Guertel 18-20, A-1090 Vienna, Austria. klaus.friedrich@meduniwien.ac.at
Purpose:
To evaluate qualitative and quantitative image quality parameters of isotropic three-dimensional (3D) cartilage-imaging magnetic resonance (MR)-sequences at 3T.
Materials And Methods:
The knees of 10 healthy volunteers (mean age, 24.4±5.6 years) were scanned at a 3T MR scanner with water-excited 3D Fast-Low Angle Shot (FLASH), True Fast Imaging with Steady-state Precession (TrueFISP), Sampling Perfection with Application-optimized Contrast using different flip-angle Evolutions (SPACE) as well as conventional and two individually weighted Double-Echo Steady-State (DESS) sequences. The MR images were evaluated qualitatively and quantitatively (signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), SNR efficiency, CNR efficiency). Quantitative parameters were compared by means of a Tukey-test and sequences were ranked according to SNR/CNR, SNR/CNR efficiency and qualitative image grading.
Results:
The highest SNR was measured for SPACE (34.0±5.6), the highest CNR/CNR efficiency (cartilage/fluid) for the individually weighted DESS (46.9±18.0/2.18±0.84). SPACE, individually weighted and conventional DESS were ranked best with respect to SNR/CNR and SNR/CNR efficiency. The DESS sequences also performed best in the qualitative evaluation. TrueFISP performed worse, FLASH worst. The individually weighted DESS sequences were generally better than the conventional DESS with the significant increase of cartilage-fluid contrast (46.9±18.0/31.9±11.4 versus 22.0±7.3) as main advantage.
Conclusion:
Individually weighted DESS is the most promising candidate; all tested sequences performed better than FLASH.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies IV: Magnetic Resonance Imaging