Comparison of two oscillometric blood pressure monitors in subjects with atrial fibrillation
Tyler S Lamb1, Amar Thakrar, Mahua Ghosh
1Department of Medicine and Cardiovascular Risk Factor Reduction Unit, University of Saskatchewan, Saskatoon, SK, Canada.
Objective:
To compare blood pressure readings obtained with two commonly used oscillometric monitors: Omron HEM 711 AC (OM) and Welch-Allyn 52000 series NIBP/oximeter (WA) with mercury sphygmomanometers (Merc) in subjects with atrial fibrillation.
Methods:
We recruited 51 hemodynamically stable subjects with atrial fibrillation. Fifty four subjects in normal sinus rhythm served as controls. Supine blood pressure readings in each arm were recorded simultaneously using one monitor and Merc. The second monitor then replaced the first and readings were repeated. Merc was then switched to the opposite arm, and both monitors retested. Apical heart rates were ascertained with a stethoscope. We used the averaged, same arm Merc readings as "gold standard".
Results:
Automated blood pressure readings were obtained in all control subjects and in all but three of those with atrial fibrillation. Both monitors, and operators, noted a difference between apical and radial/brachial pulse rates: apical-recorded: Merc 6.1 + or - 15.0; OM 5.5 + or - 13.7; WA 10.0 + or - 21.2 beats per minute. Both monitors were accurate in controls: over 90% of readings were within 10 mmHg of averaged Merc, and both achieved European Hypertension Society standards. In subjects with atrial fibrillation, about one quarter of all oscillometric readings differed from Merc by more than 10 mmHg. Both falsely high and falsely low readings occurred, some up to 30 mmHg. There was no relation between accuracy and heart rate.
Conclusions:
Single blood pressure readings, taken with oscillometric monitors in subjects with atrial fibrillation differ, often markedly, from those taken manually. Health care professionals should record multiple readings manually, using validated instruments when making therapeutic decisions.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure