Background:
Evaluation of repolarization during sequentional biventricular pacing.
Methods:
Patients with biventricular devices, and left ventricular leads placed to the basal part of lateral left ventricular wall were enrolled. QRS, QTc, JTc, and corrected Tpeak-Tend intervals were compared during sequentional biventricular, left ventricular, and right ventricular pacing.
Results:
Five patients with nonischemic and five with ischemic cardiomyopathy due to anterior myocardial infarction were enrolled. No correlation was observed between values of repolarization among patients. The optimal values of repolarization were significantly different from values of echocardiographically guided hemodynamic optimization. Two patients with biventricular pacing-induced ventricular fibrillation were successfully treated by reprogramming of V-V delay according to interventricular delay resulting in shorter Tpeak-Tend interval, although delayed effect of amiodarone in one of these patients cannot be ruled out.
Conclusions:
Patients with biventricular devices may be prone to development of ventricular arrhythmias depending on programmed V-V interval. We suggest that optimization of repolarization may be performed in patients with biventricular pacemakers in the absence of backup ICD and those with frequent episodes of ventricular tachyarrhythmias, although this finding deserves further study.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Mitral Regurgitation III: Medical Management
Heart Failure Drugs: Inotropic Agents
Dysrhythmias VI: Management of Dysrhythmias
Mitral Regurgitation IV: Nursing Management
Cardiomyopathy V: Interprofessional Care
Electrophysiology of Normal Cardiac Rhythm