System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197
Guido Freckmann1, Annette Baumstark, Nina Jendrike
1Institute for Diabetes-Technology at the University of Ulm, Germany.
Background:
Blood glucose (BG) monitoring systems enable diabetes patients to effectively control and adjust their therapy. BG monitoring systems with a Conformité Européenne (CE) label should meet the standard DIN EN ISO 15197:2003: > or =95% of the BG results shall fall within +/-15 mg/dL of the reference method at BG concentrations <75 mg/dL and within +/-20% at BG concentrations > or =75 mg/dL. We intended to verify if BG monitoring systems with a CE label fulfill these minimum accuracy requirements.
Methods:
We evaluated 27 BG monitoring systems from 18 manufacturers for system accuracy according to DIN EN ISO 15197:2003. Twenty-four systems were compared with the glucose oxidase reaction (YSI 2300 glucose analyzer [YSI Life Sciences, Yellow Springs, OH]) and three systems with the hexokinase reaction (Hitachi 917 [Roche Diagnostics GmbH, Mannheim, Germany]). Duplicate measurements of 100 blood samples with a defined distribution of BG concentrations from 20 mg/dL to 600 mg/dL from > or =100 subjects were included in the evaluation.
Results:
Sixteen of the 27 BG monitoring systems fulfilled the minimum accuracy requirements of the standard, i.e., > or =95% of their results showed the minimum acceptable accuracy. Overall, the mean percentage of results showing the minimum acceptable accuracy was 95.2 +/- 5.2%, ranging from 80.0% to 100.0%.
Conclusions:
More than 40% of the evaluated BG monitoring systems did not fulfill the minimum accuracy requirements of DIN EN ISO 15197:2003. As inaccurate BG monitoring systems bear the risk of false treatment decisions by the diabetes patient and subsequent possible severe health injury, manufacturers should regularly and effectively check the quality of BG meters and BG test strips.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure in the Leg
Preparation:
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: