Related Experiment Video
Updated: Jun 16, 2026

Measurement of Aerosols Optical Thickness of the Atmosphere using the GLOBE Handheld Sun Photometer
Published on: May 29, 2019
Point Visibility Meter: a forward scatter instrument for the measurement of aerosol extinction coefficient
Abstract:
A new instrument for the measurement of atmospheric extinction coefficient-the Point Visibility Meter (PVM)-is introduced and its principle of operation discussed. The instrument measures the amount of light scattered at a particular forward scattering angle by haze or fog; the extinction coefficient sigma is then found from the empirical law relating sigma to the scatter at this angle. Detailed calculations of aerosol scattering properties using Mie theory are given to support this empirical law. Error estimates based on these calculations indicate rms errors of less than +/-12% for a range of sigma values 2-60 km(-1). Field trials of the instrument against standard transmissometers confirm these theoretical findings and show a linearity of instrumental output with extinction coefficient, with an rms error of +/-11.7%.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
UV–Vis Spectrometers
Atomic Emission Spectroscopy: Instrumentation
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Atomic Emission Spectroscopy: Overview
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview