Optimization of initial energy for cardioversion of atrial tachyarrhythmias with biphasic shocks
Johann Reisinger1, Christine Gstrein, Thomas Winter
1Department of Internal Medicine/Cardiology, Krankenhaus Barmherzige Schwestern, Seilerstätte 4, 4010 Linz, Austria. johann.reisinger@bhs.at
Objective:
Recommendations for optimal first-shock energies with biphasic waveforms are conflicting. We evaluated prospectively the relation between type and duration of atrial tachyarrhythmias and the probability of successful cardioversion with a specific biphasic shock waveform to develop recommendations for the initial energy setting aiming at the lowest total cumulative energy with 2 or less consecutive shocks.
Methods:
We analyzed 453 consecutive patients undergoing their first transthoracic electrical cardioversion, including 358 attempts for atrial fibrillation (AF) and 95 attempts for atrial flutter (AFL) or atrial tachycardia (AT). A step-up protocol with a truncated exponential biphasic waveform starting at 50 J was used. Total cumulative energies were estimated under the assumption of a 2-tiered escalating shock protocol with different initial energy settings and a "rescue shock" of 250 J for AFL/AT or 360 J for AF. The initial energy setting leading to the lowest total cumulative energy was regarded as the optimal first-shock level.
Results:
Cardioversion was successful in 448 patients (cumulative efficacy, 99 %). In patients with AFL/AT, the lowest total cumulative energy was attained with an initial energy setting of 50 J. In patients with AF, lowest values were achieved with an initial energy of 100 J for arrhythmia durations of 2 days or less and an initial energy of 150 J for arrhythmia durations of more than 2 days.
Conclusion:
We recommend an initial energy setting of 50 J in patients with AFL/AT, of 100 J in patients with AF 2 days or less, and of 150 J with AF more than 2 days.
More Related Videos
08:43Advanced Cardiac Rhythm Management by Applying Optogenetic Multi-Site Photostimulation in Murine Hearts
Published on: August 26, 2021
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiopulmonary Resuscitation IV: Pharmacological Management
Cardiopulmonary Resuscitation III: AED Use
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias II: Classification of Tachyarrhythmias