Amperometric glucose sensors: sources of error and potential benefit of redundancy
Jessica R Castle1, W Kenneth Ward
1Oregon Health Services University, Portland, Oregon, USA.
Abstract:
Amperometric glucose sensors have advanced the care of patients with diabetes and are being studied to control insulin delivery in the research setting. However, at times, currently available sensors demonstrate suboptimal accuracy, which can result from calibration error, sensor drift, or lag. Inaccuracy can be particularly problematic in a closed-loop glycemic control system. In such a system, the use of two sensors allows selection of the more accurate sensor as the input to the controller. In our studies in subjects with type 1 diabetes, the accuracy of the better of two sensors significantly exceeded the accuracy of a single, randomly selected sensor. If an array with three or more sensors were available, it would likely allow even better accuracy with the use of voting.
Related Concept Videos
Amperometry: Overview
Microbial Biosensors
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Errors occurring during blood pressure monitoring
Several factors...
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in the...