Identification of Rare Bacterial Pathogens by 16S rRNA Gene Sequencing and MALDI-TOF MS
Published on: July 11, 2016
Accuracy of 6 commercial systems for identifying clinical Aeromonas isolates
Brigitte Lamy1, Frédéric Laurent, Isabelle Verdier
1Centre Hospitalier du Bassin de Thau, Laboratoire de Biologie, Sète, France. blamy@ch-bassindethau.fr
Abstract:
We compared the accuracy of 6 commercial systems for Aeromonas identification by testing 87 clinical isolates in routine conditions, using partial rpoB gene sequencing as the reference standard. The systems were API-20E, API-32GN, the ID-GN card with the Vitek2 system (bioMérieux, Marcy l'Etoile, France), the identification portion of the NFC47 panel (MicroScan Walk/Away system; Siemens Healthcare, Sacramento, CA), ID69 (Phoenix system; BD Diagnostic Systems, Sparks, MD), and GN2 microplates (Omnilog system; Biolog, Hayward, CA), for which 67 (77.1%), 80 (91.9%), 72 (82.7%), 70 (80.5%), 64 (73.5%), and 59 (67.8%) isolates, respectively, were correctly identified at the genus and species level. Confusion with Vibrio affected 6.9% and 16.1% of results obtained with NFC47 and API-20E, respectively. Overall, the accuracy of identification for aeromonads was hampered by outdated databases and taxonomy, weak algorithms, and impractical additional tests. Commercial identification systems should be redesigned to make Aeromonas identification algorithms more robust and to cover infrequent clinical species of this genus.
More Related Videos
09:07Direct Microbial Identification using An Automated Microbial Identification System to Facilitate the EUCAST RAST Method Without Mass Spectrometry
Published on: May 24, 2024
11:25Preparation of a Blood Culture Pellet for Rapid Bacterial Identification and Antibiotic Susceptibility Testing
Published on: October 15, 2014
Related Concept Videos
Automated Microbial Diagnostics
Rapid Identification of Pathogens
Methods of Classification and Identification