Accuracy of different methods for blood glucose measurement in critically ill patients
Sérgio Antônio Pulzi Júnior1, Murillo Santucci Cesar de Assunção, Bruno Franco Mazza
1Intensive Care Sector, Escola Paulista de Medicina, Universidade Federal de São Paulo, São Paulo, Brazil.
Context And Objective:
Although glucometers have not been validated for intensive care units, they are regularly used. The aim of this study was to compare and assess the accuracy and clinical agreement of arterial glucose concentration obtained using colorimetry (Agluc-lab), capillary (Cgluc-strip) and arterial (Agluc-strip) glucose concentration obtained using glucometry and central venous glucose concentration obtained using colorimetry (Vgluc-lab).
Design And Setting:
Cross-sectional study in a university hospital.
Method:
Forty patients with septic shock and stable individuals without infection were included. The correlations between measurements were assessed both in the full sample and in subgroups using noradrenalin and presenting signs of tissue hypoperfusion.
Results:
Cgluc-strip showed the poorest correlation (r = 0.8289) and agreement (-9.87 +/- 31.76). It exceeded the limits of acceptable variation of the Clinical and Laboratory Standards Institute in 23.7% of the cases, and was higher than Agluc-lab in 90% of the measurements. Agluc-strip showed the best correlation (r = 0.9406), with agreement of -6.75 +/- 19.07 and significant variation in 7.9%. For Vgluc-lab, r = 0.8549, with agreement of -4.20 +/- 28.37 and significant variation in 15.7%. Significant variation was more frequent in patients on noradrenalin (36.4% versus 6.3%; P = 0.03) but not in the subgroup with hypoperfusion. There was discordance regarding clinical management in 25%, 22% and 15% of the cases for Cgluc-strip, Vgluc-lab and Agluc-strip, respectively.
Conclusion:
Cgluc-strip should be avoided, particularly if noradrenalin is being used. This method usually overestimates the true glucose levels and gives rise to management errors.
Clinical Trial Registration:
ACTRN12608000513314 (registered as an observational, cross-sectional study).
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Hypoglycemia and Glucagon
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Hyperglycemia
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.