Related Experiment Video
Updated: Jun 16, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Implementation of a new low cost detector for low intensity light pulses produced by a radioactive gamma-source
V Bellini1, M Capogni, V Febbraro
1INFN Sezione di Catania, Catania, Italy.
Abstract:
The readout of a cheap scintillating fibre was implemented by means of a new solid-state device (Silicon Photomultiplier, SiPM), able to detect very tiny light pulses. The GEANT3.21 code was adopted to simulate the counting detection efficiency and the energy deposited inside the scintillating fibre due to the radiation of a radioactive gamma source. The results obtained show the capability to detect gamma rays producing as few as 3-4 photoelectrons. The new devices may have many applications in radioactivity metrology.
Related Concept Videos
Determination of Crystal Structures
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...