Related Experiment Video
Updated: Jun 16, 2026

Picometer-Precision Atomic Position Tracking through Electron Microscopy
Published on: July 3, 2021
Calibration of electron spectrometer resolution in attosecond streak camera
Ximao Feng1, Steve Gilbertson, Sabih D Khan
1J. R. Macdonald Laboratory, Department of Physics, Kansas State University, Manhattan, KS 66506, USA.
Abstract:
We report a new method for determining the energy resolution of time-of-flight spectrometers for detecting photoelectrons produced with attosecond XUV pulses. By measuring the width of the 2s2p autoionization line of helium, we found the resolution of our spectrometer to be approximately 0.6 eV for electrons at 35.5 eV. Furthermore, the resolution in the 10 to 35 eV range was determined by applying a retarding potential at the entrance of the drift tube.
Related Concept Videos
Atomic Emission Spectroscopy: Instrumentation
Mass Analyzers: Overview
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Overview