Evaluation of the In2it analyzer for HbA1c determination
1Laboratory of Pediatric Biology and Research, American Memorial Hospital, CHU Reims, 47, rue Cognacq-Jay, 51092 Reims cedex, France.
Diabetes & Metabolism
|February 26, 2010
Summary
The In2it analyzer offers a rapid and reliable method for measuring hemoglobin A1c (HbA1c), suitable for both laboratory and point-of-care use. While analytical performance is good, software improvements are needed for optimal application in diabetes management.
Area of Science:
- Clinical Chemistry
- Analytical Biochemistry
- Diabetes Mellitus Diagnostics
Background:
- Hemoglobin A1c (HbA1c) is crucial for monitoring glycemic control in diabetes.
- Accurate and reliable HbA1c assays are essential for effective patient management.
- New analytical devices are needed to meet the demands of clinical settings.
Purpose of the Study:
- To evaluate the analytical performance and practicability of the new In2it analyzer for HbA1c determination.
- To assess the In2it analyzer's suitability for both laboratory and point-of-care testing.
- To compare the In2it analyzer with established HPLC and immunoassay methods.
Main Methods:
- Analytical performance testing of the In2it analyzer, including precision, interference, and linearity.
- Correlation studies with a high-performance liquid chromatography (HPLC)-based Variant II analyzer and an immunoassay-based DCA 2000 analyzer.
- Evaluation of the analyzer's practicability and compliance with good laboratory practices.
Main Results:
- The In2it analyzer demonstrated good linearity (4-14% HbA1c) and precision (CVs 2.4-3.9%).
- Satisfactory correlation was observed with HPLC Variant II (r²=0.974) and DCA 2000 (r²=0.794), though some random differences occurred with DCA 2000.
- The method showed no significant interference from common hemoglobin variants, labile HbA1c, or common clinical conditions like hyperbilirubinaemia and hypertriglyceridaemia.
Conclusions:
- The In2it analyzer is suitable for HbA1c testing in small laboratory batches and point-of-care settings.
- Overall analytical performance is satisfactory, but software requires upgrades for quality control, traceability, and data security.
- Caution is advised regarding potential underestimation of HbA1c results in cases of high hypertriglyceridaemia.