K A Brennan1, G Eapen, D Turnbull
1Department of Anaesthesia, Sheffield Teaching Hospitals NHS Trust, Royal Hallamshire Hospital, UK. karlbrennan@talktalk.net
Background:
In 2008, the National Patient Safety Agency (NPSA) published a report after 42 incidents and two deaths where glucose-containing flush solutions were attached to the arterial line. The molar concentration of 5% glucose is 277 mmol litre(-1). Only a tiny amount of sample contamination will lead to an artificially high glucose. As the NPSA sought a solution, a bench model was constructed to compare the performance of three open and three closed arterial line systems in limiting sample contamination.
Methods:
All arterial line systems were set up in a standard manner and pressurized to 300 mm Hg with 5% glucose used as the flush solution. This was connected to the 'radial artery' using an 18 G needle representing the radial cannula. The radial artery was simulated using a wide-bore extension set with 'blood' flow at 60 ml min(-1). Blood was simulated by the addition of red dye to Hartmann's solution. Increasing multiples of arterial line dead space were aspirated and discarded. Blood samples were then obtained and glucose concentration was measured.
Results:
Significant glucose contamination (3 mmol litre(-1) +/-3.4) was detected in all open arterial line systems up to an aspiration volume of five times the dead space. No samples from the closed systems recorded glucose concentration >1 mmol litre(-1).
Conclusions:
Recommended minimal discard volumes are inadequate in the presence of glucose as the flush solution and can lead to high blood glucose readings, inappropriate insulin use, and iatrogenic neuroglycopaenia. Our study demonstrates that the closed-loop arterial sampling system could be the universal solution sought by the NPSA.
Related Concept Videos
Hypoglycemia and Glucagon
Hypoglycemia
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
Hyperglycemia
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Oral Hypoglycemic Agents: Glinides