Respiratory distress associated with inadequate mechanical ventilator flow response in a neonate with congenital
Brian K Walsh1, Nancy Craig, Peter Betit
1Respiratory Care Department, Children's Hospital Boston, Boston, Massachusetts 02115, USA. brian.walsh@childrens.harvard.edu
Abstract:
The incidence of congenital diaphragmatic hernia has been reported as 0.17-0.66 per 1,000 births. Despite advances in neonatal intensive care, congenital diaphragmatic hernia is associated with high mortality and morbidity. We report a neonate who was born with a left congenital diaphragmatic hernia and underwent surgical repair. The lack of ventilator flow response and flow cycling was identified via interpretation of the ventilator graphic and clinical assessment. Presumably, the ventilator failed to respond to the patient's peak inspiratory flow demand, despite the clinician's setting the highest peak flow available. A time-cycled pressure-limited mode with adjustable peak flow rate was the only option that met the infant's flow requirement, and alleviated the respiratory distress. This clinical finding follows bench research that raises the concern that so called "cradle-to-grave" ventilators may not optimally support all neonates.
Related Concept Videos
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Pulmonary Cycle: Exhalation
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Acute Respiratory Failure-IV
Acute Respiratory Failure-III
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)