A bench test to confirm the core features of volume-assured non-invasive ventilation
Nicholas S Oscroft1, Ian E Smith
1Respiratory Support and Sleep Centre, Papworth Hospital NHS Foundation Trust, Papworth Everard, Cambridge, UK. nickoscroft@doctors.net.uk
Unlabelled:
Volume-assured non-invasive ventilation (NIV) theoretically guarantees minute ventilation with circuit leak compensation unlike other modes of NIV. Bench testing demonstrated that minute ventilation was maintained with varying lung compliance and resistance with minimal effect from circuit leak, confirming for the first time the core features of volume-assured NIV.
Background And Objective:
Volume-assured non-invasive positive pressure ventilation (va-NIPPV) is a novel mode designed to adapt pressure support (PS) to achieve a target minute ventilation (TgV). This may optimize ventilation; however, no data confirm that va-NIPPV compensates appropriately for the changes in pulmonary mechanics and circuit leak seen in clinical practice. Bench testing assessed these principles.
Methods:
A ventilator featuring a va-NIPPV mode was studied. A test lung with varying compliance and resistance, and pneumotachograph were used. Eight lung model settings were chosen: (i) low resistance and high compliance; (ii) low resistance and low compliance; (iii) high resistance and high compliance; and (iv) high resistance and low compliance, all with and without additional circuit leak. An expiration valve, respiratory rate of 15, inspiratory time of 1 s and PS between 3 and 21 cm H2O were used. Va-NIPPV was tested with varying TgV after establishing the range of minute ventilation possible in a pressure preset mode.
Results:
At a TgV of 10 L/min, va-NIPPV delivered minute ventilation of (median (interquartile range) ): 11 (10.9-11, 10.2 (10.2-10.3), 12.4 (12.4-12.4) and 11.2 (10.9-11.2) L/min in test lung settings 1, 2, 3 and 4, respectively. Additional leak between 8-33 L/min had little effect. Similar results were seen at other TgV, within the ventilator's PS capabilities.
Conclusions:
These data confirm that va-NIPPV is able to approximate a preset TgV with varying lung compliance and resistance, and that additional circuit leak has little effect on the delivered minute ventilation.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation I: Indication and Settings
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...