Related Experiment Video
Updated: Jun 15, 2026

Fabrication of High Contact-Density, Flat-Interface Nerve Electrodes for Recording and Stimulation Applications
Published on: October 4, 2016
Selective stimulation of the human femoral nerve with a flat interface nerve electrode
M A Schiefer1, K H Polasek, R J Triolo
1Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, USA. matthew.schiefer@case.edu
Abstract:
In humans, we tested the hypothesis that a flat interface nerve electrode (FINE) placed around the femoral nerve trunk can selectively stimulate each muscle the nerve innervates. In a series of intraoperative trials during routine vascular surgeries, an eight-contact FINE was placed around the femoral nerve between the inguinal ligament and the first nerve branching point. The capability of the FINE to selectively recruit muscles innervated by the femoral nerve was assessed with electromyograms (EMGs) of the twitch responses to electrical stimulation. At least four of the six muscles innervated by the femoral nerve were independently and selectively recruited in all subjects. Of these, at least one muscle was a hip flexor and at least two were knee extensors. Results from the intraoperative experiments were used to estimate the potential for the electrode to restore knee extension and hip flexion through functional electrical stimulation. Normalized EMGs and biomechanical simulations were used to estimate joint moments and functional efficacy. Estimated knee extension moments exceed the threshold required for the sit-to-stand transition.

