Venous needle dislodgement: the pitfalls of venous pressure measurement and possible alternatives, a review
1Medical Devices Consultant, Koestenberg, Austria. hdp@aon.at
Abstract:
The historical development of the use of pressure sensors as protective system against blood loss to the environment is reviewed. In spite of early warnings about the deficiency of such sensors, the venous pressure sensor is still formally accepted as protective system against blood loss to the environment in case of leaks or needle dislodgement. The early warnings were corroborated by publications and accident reports. Several alternative methods have been developed or described in the literature recently. These methods are critically reviewed. The conclusion is that external monitors are currently the only available alternatives, although long-time clinical experience is missing. Methods employing sensors integrated into dialysis machines have been described but it is unlikely that any of these methods will become available in the immediate future.
More Related Videos
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...
Assessing Blood pressure in the Leg
Preparation:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...