Automated Microbial Diagnostics01:24

Automated Microbial Diagnostics

Automated diagnostic analyzers have transformed clinical microbiology by providing rapid and reliable methods for pathogen identification and antibiotic susceptibility testing. Among these systems, the Vitek 2 is widely used because it automates the traditionally labor-intensive processes of microbial identification (ID) and antibiotic susceptibility testing (AST), delivering standardized and timely results that are essential for effective patient care.Microbial Identification with ID CardsThe...

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Reply to McArdle and Austin, "Further detail required to assess additional evidence on agreement between VITEK 2 and broth microdilution piperacillin/tazobactam testing in cephalosporin resistant Enterobacterales".

Journal of clinical microbiology·2026
Same author

Viewpoint: global antimicrobial stewardship accreditation is needed: the experience in Kentucky, USA.

The Journal of antimicrobial chemotherapy·2025
Same author

Effect of a rapid bacteraemia response programme on <i>Staphylococcus aureus</i> bacteraemia management and mortality.

JAC-antimicrobial resistance·2025
Same author

A network analysis of carbapenem-resistant Klebsiella pneumoniae among healthcare facilities.

Scientific reports·2025
Same author

Evaluation of MicroScan and VITEK 2 systems for susceptibility testing of Enterobacterales with updated breakpoints.

Journal of clinical microbiology·2025
Same author

The American Society for Microbiology collaboration with the CDC Laboratory Medicine Best Practices initiative for evidence-based laboratory medicine.

Clinical microbiology reviews·2024

Related Experiment Video

Updated: Jun 15, 2026

A Semi-automated Approach to Preparing Antibody Cocktails for Immunophenotypic Analysis of Human Peripheral Blood
08:17

A Semi-automated Approach to Preparing Antibody Cocktails for Immunophenotypic Analysis of Human Peripheral Blood

Published on: February 8, 2016

Comparison of BD Phoenix AP Workflow with Vitek 2.

Alan D Junkins1, Sophie S Arbefeville, Wanita J Howard

  • 1Department of Pathology, C606 GH, University of Iowa, Roy J. and Lucille A. Carver College of Medicine, 200 Hawkins Drive, Iowa City, IA 52242-1009, USA.

Journal of Clinical Microbiology
|March 12, 2010
PubMed
Summary

The BD Phoenix AP instrument significantly reduced manual setup time by 50% compared to Vitek 2. This automated system offers faster processing for microbial identification and antibiotic susceptibility testing.

More Related Videos

Automating Citrus Budwood Processing for Downstream Pathogen Detection Through Instrument Engineering
11:30

Automating Citrus Budwood Processing for Downstream Pathogen Detection Through Instrument Engineering

Published on: April 21, 2023

Related Experiment Videos

Last Updated: Jun 15, 2026

A Semi-automated Approach to Preparing Antibody Cocktails for Immunophenotypic Analysis of Human Peripheral Blood
08:17

A Semi-automated Approach to Preparing Antibody Cocktails for Immunophenotypic Analysis of Human Peripheral Blood

Published on: February 8, 2016

Automating Citrus Budwood Processing for Downstream Pathogen Detection Through Instrument Engineering
11:30

Automating Citrus Budwood Processing for Downstream Pathogen Detection Through Instrument Engineering

Published on: April 21, 2023

Area of Science:

  • Clinical Microbiology
  • Medical Technology
  • Laboratory Automation

Background:

  • Manual setup in microbial identification and antibiotic susceptibility testing (AST) can be time-consuming.
  • Automation aims to improve laboratory efficiency and reduce hands-on time.
  • The BD Phoenix system is a widely used platform for microbial analysis.

Purpose of the Study:

  • To evaluate the impact of the BD Phoenix AP instrument on manual setup time.
  • To compare the efficiency of the BD Phoenix system with the AP instrument against the Vitek 2 system.

Main Methods:

  • A comparative study was conducted using the BD Phoenix system with and without the AP instrument.
  • Manual manipulation time per isolate was recorded for batches of 14 organisms.
  • Time data was compared between the BD Phoenix system (with AP instrument) and the Vitek 2 system.

Main Results:

  • The BD Phoenix AP instrument reduced manual setup time by 50%.
  • Average manual manipulation time per isolate was 89.5 seconds for BD Phoenix with AP, versus 101 seconds for Vitek 2 (P<0.001).

Conclusions:

  • The BD Phoenix AP instrument significantly decreases manual manipulation time in microbial analysis.
  • This automation enhances the efficiency of the BD Phoenix system for both identification and AST.
  • The AP instrument offers a more time-efficient workflow compared to Vitek 2.