High-flow nasal cannula and extubation success in the premature infant: a comparison of two modalities
1Division of Newborn Medicine, Weill Cornell Medical College, New York, NY, USA.
Insights
This pilot study found no significant difference in extubation success rates between Fisher and Paykel (FP) and Vapotherm (VT) high-flow nasal cannula (HFNC) for premature infants. Both devices showed similar efficacy in preventing reintubation in this small cohort.
Area of Science:
- Neonatal Medicine
- Respiratory Care
- Clinical Trials
Background:
- Premature infants often require respiratory support after extubation.
- High-flow nasal cannula (HFNC) is a common method for post-extubation support.
- Comparing different HFNC devices is crucial for optimizing care.
Purpose of the Study:
- To compare the effectiveness of Fisher and Paykel (FP) and Vapotherm (VT) HFNC devices.
- To evaluate the prevention of reintubation within 72 hours and 7 days post-extubation.
- To assess extubation success in premature infants (26-29 weeks gestation).
Main Methods:
- Prospective, randomized pilot study design.
- Involved 40 premature infants (26-29 weeks gestation).
- Compared FP HFNC versus VT HFNC for post-extubation support.
Main Results:
- Extubation failure rate at 72 hours was 18% for FP and 9% for VT.
- Reintubation rate within 7 days was 30% for FP and 27% for VT.
- No statistically significant differences were observed between the two HFNC devices.
Conclusions:
- Fisher and Paykel (FP) and Vapotherm (VT) HFNC devices demonstrated comparable effectiveness.
- No significant difference in extubation success rates between FP and VT HFNC was found.
- Further research with larger sample sizes may be warranted.
Objective:
To compare the effectiveness of Fisher and Paykel (FP) and Vapotherm (VT) high-flow nasal cannula (HFNC) in preventing reintubation either within 72 h or <7 days after extubation of premature infants. The primary outcome was the rate of extubation failure defined as reintubation within 72 h. Secondary outcomes included reintubation after ≤7 days.
Study Design:
This was a prospective, randomized pilot study comparing the extubation success of 40 infants born between 26 and 29 weeks of gestation.
Result:
The rate of extubation failure at 72 h was 18% for FP and 9% for VT. The failure rate ≤7 days after extubation was 30% for FP and 27% for VT. None of these differences were statistically significant.
Conclusion:
There was no difference between FP and VT in the extubation success of infants born between 26 and 29 weeks.
Related Concept Videos
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Suctioning the Nasopharyngeal Airway
Equipment Required
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...