Raman and IR Spectroelectrochemical Methods as Tools to Analyze Conjugated Organic Compounds
Published on: October 12, 2018
[Raman spectra of carbon fibers during electrochemical treatment]
Min Zhang1, Bo Zhu, Cheng-guo Wang
1Key Laboratory of Liquid Structure and Heredity of Materials, Ministry of Education, Shandong University, Ji' nan 250061, China. girlzm05sd@gmail.com
Abstract:
Laser Raman spectroscopy was employed to characterize the microstructure variations of polyacrylonitrile-based carbon fibers during electrochemical treatment, and the characteristics of first-order Raman spectra of carbon fibers with different treatment time were investigated in the present paper. The results indicate that the Raman spectra of the carbon fibers can be fitted into four bands, named as D (or D1) band, G band, D2 band and D3 band, respectively. The Raman parameters to characterize surface microstructure variations of carbon fibers mainly include R(I(D2)) / I(G), area ratio of D band and G band), I(D2) / I(G) (area ratio of D2 band and G band), I(D3) / I(G) (area ratio of D3 band and G band), and I(D(S))/ I(G) (area ratio of all the disordered structure and G band). The peak separation between D band and G band becomes large after electrochemical treatment. R increases, which indicates that the surface disordered degree of carbon fibers increases. I(D3) / I(G) increases, which is caused by organic molecules, fragments or functional groups; decreases which is caused by the break of the aliphatic structures. With increasing treatment time, I(D(S)) / I(G) increases continuously, and the change trend of l(D(S)) / I(G) is consistent with that of R value, which can be used to comprehensively explain the variation of the surface structure of carbon fibers. So, the variety rules of the structure of carbon fibers can be investigated by laser Raman spectroscopy during electrochemical treatment.
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...