Frederic Schweizer1, Gunter Christ
1Medizinische Physik, Klinik für Radioonkologie, Universitätsklinikum Tübingen, Tübingen.
Abstract:
Oblique incidence of electron beams is a rare clinical application which can be used e.g. at the irradiation of the internal mammary lymph nodes in the case of mamma carcinoma. 3-D treatment planning systems are accepted standard for photon beams but not for electron beams. The investment for measuring, implementing and testing basic data seems many customers not to be worthwhile when considering the well known inaccuracies of conventional algorithms. From Monte-Carlo (MC) based algorithms however higher accuracies can be expected. It was the aim of this paper to test the MC algorithm of the treatment planning system Oncentra Treatment Planning (OTP, Theranostic) in the application for oblique incidence of electron beams. Measured and simulated values are compared for electron energies from 4 MeV up to 18 MeV. For standard electron applicators, even at extended source-surface distances, satisfying results are achieved. However, the agreement gets lost with increasing oblique incidence, especially for small individual and asymmetrical cut outs. We therefore still use measured values instead of simulated dose distributions.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021