Study on the response of thermoluminescent dosemeters to synchrotron radiation: experimental method and Monte Carlo
A K Bakshi1, S Chatterjee, T Palani Selvam
1Health, Safety and Environment Group, Radiological Physics and Advisory Division, Bhabha Atomic Research Centre, Trombay, Mumbai 400 085, India. akbakshi123@yahoo.com
Abstract:
In the present study, the energy dependence of response of some popular thermoluminescent dosemeters (TLDs) have been investigated such as LiF:Mg,Ti, LiF:Mg,Cu,P and CaSO(4):Dy to synchrotron radiation in the energy range of 10-34 keV. The study utilised experimental, Monte Carlo and analytical methods. The Monte Carlo calculations were based on the EGSnrc and FLUKA codes. The calculated energy response of all the TLDs using the EGSnrc and FLUKA codes shows excellent agreement with each other. The analytically calculated response shows good agreement with the Monte Carlo calculated response in the low-energy region. In the case of CaSO(4):Dy, the Monte Carlo-calculated energy response is smaller by a factor of 3 at all energies in comparison with the experimental response when polytetrafluoroethylene (PTFE) (75 % by wt) is included in the Monte Carlo calculations. When PTFE is ignored in the Monte Carlo calculations, the difference between the calculated and experimental response decreases (both responses are comparable >25 keV). For the LiF-based TLDs, the Monte Carlo-based response shows reasonable agreement with the experimental response.