Related Experiment Video
Updated: Jun 14, 2026

A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
Optimization and validation of an RP-HPLC method for analysis of hydrocortisone acetate and lidocaine in
Biljana Jancic-Stojanović1, Andjelija Malenović, Slavko Marković
1University of Belgrade, Faculty of Pharmacy, Department of Drug Analysis, Vojvode Stepe 450, 11 152 Belgrade, Serbia. jancic.stojanovic@pharmacy.bg.ac.yu
Abstract:
An RP-HPLC method has been optimized and validated for the simultaneous determination of hydrocortisone acetate and of lidocaine in suppositories. For the method optimization, response surface methodology was applied, and the obtained model was tested using analysis of variance. The optimal separations were conducted on a Beckman-Coulter 150 x 4.6 mm, 5 microm particle-size column at 20 degrees C. The mobile phase was methanol-water (65 + 35, v/v), pH adjusted to 2.5 with 85% orthophosphoric acid, with a flow rate of 1.0 ml/min. UV detection was performed at 250 nm. Phenobarbital was used as an internal standard. The method was validated for selectivity, linearity, precision, and robustness.
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process: