Related Experiment Video
Updated: Jun 14, 2026

Determining Viral Disinfection Efficacy of Hot Water Laundering
Published on: June 21, 2022
Re-evaluation of current A0 value recommendations for thermal disinfection of reusable human waste containers based
M Diab-Elschahawi1, U Fürnkranz, A Blacky
1Clinical Institute for Hygiene and Medical Microbiology, Division for Hospital Hygiene, Vienna General Hospital, Medical University of Vienna, Waehringer Guertel 18-20, 1090 Vienna, Austria. magda.diab-elschahawi@meduniwien.ac.at
Abstract:
Issues over the correct cleaning and disinfection of reusable medical devices are of great concern for local infection control teams. We investigated the heat resistance of two important micro-organisms implicated in nosocomial infections, Enterococcus spp. and spores of Bacillus subtilis (a surrogate for Clostridium difficile). Disinfection with moist heat, based on the A(0) concept (EN ISO 15883-1), is the most common method for disinfection of human waste containers in the hospital setting. The aim of this study was to scrutinise the A(0) concept regarding the inactivation of the above-mentioned micro-organisms and to look for possible interferences. Experiments were carried out in phosphate-buffered saline, artificial soil in suspension (ASS) and dried artificial soil (AS). Artificial soil was used to simulate insufficient cleaning of bedpans. Micro-organisms were treated at different temperatures and time periods. It was shown that soil protected the micro-organisms against heat and that dried soil enhanced this effect, probably as a result of thermal insulation. The results of the current study do not support the general recommendation to use an A(0) of 60 for the disinfection of bedpans but warrant differential strategies depending on the expected microbial load. We generally recommend for disinfection processes aimed at vegetative bacteria an A(0) of >or=180. Furthermore we want to emphasise the utmost importance of not allowing freshly contaminated bedpans to dry for longer time periods, as the significance of thorough cleaning of contaminated human waste containers before being exposed to heat treatment for their disinfection was clearly demonstrated.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Biological Treatment of Effluent and Waste Water
Microbial Wastewater Treatment