Initial experience with the Evolution mechanical dilator sheath for lead extraction: safety and efficacy
Ayman A Hussein1, Bruce L Wilkoff, David O Martin
1Cleveland Clinic, Cleveland, Ohio, USA.
Heart Rhythm
|March 30, 2010
Summary
The Evolution mechanical dilator sheath offers a new approach for extracting pacemaker and implantable cardioverter-defibrillator (ICD) leads. Initial experience shows high success rates and no complications, suggesting it
Area of Science:
- Cardiovascular Medicine
- Medical Devices
- Electrophysiology
Background:
- The Evolution mechanical dilator sheath is a novel tool designed for cardiac lead extraction.
- It features a rotational mechanism and bladed tip to address fibrosis, a common challenge in lead removal.
Purpose of the Study:
- To evaluate the initial clinical experience and efficacy of the Evolution mechanical dilator sheath system.
- To assess success rates and complication profiles associated with its use in lead extraction procedures.
Main Methods:
- The Evolution sheath was utilized for pacemaker and implantable cardioverter-defibrillator (ICD) lead extraction in 29 patients (41 leads) between March 2008 and September 2009.
- Success and complications were assessed based on the Heart Rhythm Society expert consensus criteria.
Main Results:
- The system was used for 41 leads (20 for infection, 9 for malfunction), with a median implantation time of 65 months.
- Overall, the Evolution system demonstrated an 86% success rate in patients (33 leads), achieving 100% clinical success with no reported complications.
- While successful alone in 77% of rescue cases, it required adjunctive tools in 24% of patients.
Conclusions:
- The Evolution mechanical dilator sheath appears to be a valuable addition to the armamentarium for cardiac lead extraction.
- Preliminary findings indicate its utility and safety in managing complex lead extraction scenarios.