Differences between arterial and expired pump carbon dioxide during robotic cardiac surgery
Jun-Gol Song1, Eun Ho Lee, Dae Kee Choi
1Department of Anesthesiology and Pain Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
Expired pump carbon dioxide (PepCO(2)) is not a reliable monitor for arterial carbon dioxide (PaCO(2)) during cardiopulmonary bypass in robotic cardiac surgery. This study found significant discrepancies, indicating it
Area of Science:
- Cardiovascular Surgery
- Anesthesiology
- Medical Device Monitoring
Background:
- Accurate monitoring of arterial carbon dioxide (PaCO(2)) is crucial during cardiopulmonary bypass (CPB).
- Expired pump carbon dioxide (PepCO(2)) has been proposed as a non-invasive alternative monitoring method.
- Robotic cardiac surgery presents unique physiological challenges during CPB.
Purpose of the Study:
- To evaluate the efficacy of PepCO(2) as a continuous monitor for PaCO(2) during CPB.
- To compare PepCO(2) monitoring with standard PaCO(2) measurements in patients undergoing robotic cardiac surgery.
- To assess the reliability of PepCO(2) across different phases of CPB (cooling, hypothermia, rewarming).
Main Methods:
- Prospective control study conducted in a single university institution.
- Sixty patients undergoing cardiac surgery were included.
- PepCO(2) was measured from the oxygenator exhaust, alongside simultaneous arterial blood gas analysis for PaCO(2) during CPB.
Main Results:
- Significant differences were observed between temperature-uncorrected PaCO(2) and PepCO(2) during cooling and stable hypothermia (p < 0.001).
- No significant differences were found between temperature-corrected PaCO(2) and PepCO(2) during cooling and rewarming phases (p > 0.122).
- Significant discrepancies highlight temperature-dependent variations and limitations of PepCO(2) monitoring.
Conclusions:
- Expired pump carbon dioxide (PepCO(2)) is not suitable for continuous monitoring of arterial carbon dioxide (PaCO(2)) during CPB in robotic cardiac surgery.
- The observed discrepancies, particularly during cooling and hypothermia, limit its clinical utility.
- Further research may be needed to explore alternative non-invasive monitoring techniques for CO(2) during CPB.
More Related Videos
08:12Invasive Hemodynamic Monitoring of Aortic and Pulmonary Artery Hemodynamics in a Large Animal Model of ARDS
Published on: November 26, 2018
05:25Surgical Porcine Model of Chronic Myocardial Ischemia Treated by Exosome-laden Collagen Patch and Off-pump Coronary Artery Bypass Graft
Published on: September 15, 2023