Alexander H Slocum1, William R Bosworth, Anirban Mazumdar
1Massachusetts Institute of Technology, Department of Mechanical Engineering, Cambridge, MA 02139.
Abstract:
This paper focuses on the design and implementation of a percutaneous catheter-based device to provide physicians with an externally controlled tool capable of manipulating and cutting specific chordae tendinae within the hear to alleviate problems associated with some forms of mitral valve regurgitationt. In the United States alone, approximately 500,000 people develop ischemic or functional MR per year, and the chordae tendinae cutting procedure and device are needed because many patients do not have the required level of health necessary to survive open-heart surgery. A deterministic design process was used to generate several design concepts and then evaluate and compare each concept based on a set of functional requirements. A final concept to be alpha prototyped was then chosen, further developed, and fabricated. Experiments showed that the design was capable of locating and grabbing a chord and that ultrasound imaging is a viable method for navigating the device inside of the human body. Once contact between the chord and an RF ablator tip was confirmed, the chord was successfully ablated.