Related Experiment Video
Updated: Jun 14, 2026

Purification and Analytics of a Monoclonal Antibody from Chinese Hamster Ovary Cells Using an Automated Microbioreactor System
Published on: May 1, 2019
A reversed-phase high-performance liquid chromatography method for bovine serum albumin assay in pharmaceutical
1Department of Pharmaceutics, Faculty of Pharmacy, Shiraz University of Medical Sciences, PO Box 71345-1583, Shiraz, Iran. hamidim@zums.ac.ir
Abstract:
Bovine serum albumin (BSA) is among the most widely used proteins in protein formulations as well as in the development of novel delivery systems as a typical model for therapeutic/diagnostic proteins and the new versions of vaccines. The development of reliable and easily available assay methods for quantitation of this protein would therefore play a crucial role in these types of studies. A simple gradient reversed-phase high-performance liquid chromatography with ultra-violet detection (HPLC-UV) method has been developed for quantitation of BSA in dosage forms and protein delivery systems. The method produced linear responses throughout the wide BSA concentration range of 1 to 100 micro g/mL. The average within-run and between-run variations of the method within the linear concentration range of BSA were 2.46% and 2.20%, respectively, with accuracies of 104.49% and 104.58% for within-run and between-run samples, respectively. The limits of detection (LOD) and quantitation (LOQ) of the method were 0.5 and 1 microg/mL, respectively. The method showed acceptable system suitability indices, which enabled us to use it successfully during our particulate vaccine delivery research project.
Related Concept Videos
Western Blotting
The technique begins with separating proteins from the sample using sodium dodecyl sulfate-polyacrylamide gel electrophoresis (SDS-PAGE), followed by protein transfer, immunoblotting, and finally, protein detection.
Affinity Chromatography