The intrauterine probe electrode
T H Strong1, R H Paul, G D Park
1Department of Obstetrics and Gynecology, University of Southern California School of Medicine, Los Angeles.
An intrauterine probe electrode was inserted into 100 laboring women and 366 bipolar electrode combinations were tested. A noise-cancelling technique was used with the final 28 subjects to remove competing maternal cardiac signals. Twenty-four (86%) had fetal heart rate tracings with sufficient technical quality to allow determination of the baseline fetal heart rate.
An intrauterine probe electrode was inserted into 100 laboring women and 366 bipolar electrode combinations were tested. A noise-cancelling technique was used with the final 28 subjects to remove competing maternal cardiac signals. Twenty-four (86%) had fetal heart rate tracings with sufficient technical quality to allow determination of the baseline fetal heart rate.