Related Experiment Video
Updated: Jun 14, 2026

Improving IV Insulin Administration in a Community Hospital
Published on: June 11, 2012
Capillary glucose meter accuracy and sources of error in the ambulatory setting
Helen Lunt1, Christopher Florkowski, Michael Bignall
1Christchurch Diabetes Centre and Department of Medicine, University of Otago, Christchurch, New Zealand. helen.lunt@cdhb.govt.nz
Abstract:
Hand-held glucose meters are used throughout the health system by both patients with diabetes and also by health care practitioners. Glucose meter technology is constantly evolving. The current generation of meters and strips are quick to use and require a very small volume of blood. This review aims to describe meters currently available in New Zealand, for use in the ambulatory setting. It also aims to discuss the limits of meter performance and provide technical information that is relevant to the clinician, using locally available data. Commoner causes and consequences of end-user (patient and health professional) error are illustrated using clinical case examples. No meter offers definite advantages over other meters in all clinical situations, rather meters should be chosen because they fit the needs of individual patients and because the provider is able to offer appropriate educational and quality assurance backup to the meter user. A broad understanding of the advantages and disadvantages of the subsidised meter systems available in New Zealand will help the health practitioner decide when it is in the best interests of their patients to change or update meter technology.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Uncertainty in Measurement: Accuracy and Precision
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Types of Errors: Detection and Minimization
Absolute error in a measurement is the numerical difference from the true or central value. Relative error is the ratio between absolute error and the true or central value, expressed as a percentage.
Errors can be classified by source, magnitude, and sign. There are three types of errors: systematic, random, and gross.
Systematic or...
Hyperglycemia
Uncertainty in Measurement: Reading Instruments