Related Experiment Video
Updated: Jun 14, 2026

Dynamic Visual Tests to Identify and Quantify Visual Damage and Repair Following Demyelination in Optic Neuritis Patients
Published on: April 14, 2014
Correlation between optics quality of multifocal intraocular lenses and visual acuity: tolerance to modulation
Adelina Felipe1, Francisco Pastor, José M Artigas
1Fundación Oftalmológica del Mediterráneo, Departamento de Optica, Facultad de Física, Universitat de Valencia, Burjassot, Valencia, Spain. adelina.felipe@uv.es
Purpose:
To study the relationship between the optics quality of multifocal intraocular lenses (IOLs) and distance-corrected near, intermediate, and distance visual acuity to determine the degree to which optics quality must change to produce a significant difference in visual acuity.
Setting:
Fundación Oftalmológica del Mediterráneo, Valencia, Spain.
Methods:
Monocular distance-corrected visual acuity (decimal and logMAR) was measured at 4 m, 70 cm, and 30 cm under photopic and mesopic conditions in eyes with 1 of 3 multifocal IOL models. Visual acuity versus the "average modulation" of the IOL was assessed.
Results:
The pupil diameter and patient age were similar between the 3 IOL groups (20 eyes each). A difference in average modulation (ie, optics quality) up to 15% did not produce significant differences in mean visual acuity between 2 groups; however, there was a significant difference in mean visual acuity between 2 groups when the average modulation values differed by at least 25%. The slope of the linear correlation between visual acuity and average modulation was 0.018 (r(2) = 0.91) under photopic conditions and 0.024 (r(2) = 0.089) under mesopic conditions.
Conclusions:
There was a strong linear correlation between IOL optics quality and visual acuity with distance correction under photopic conditions with a pupil of approximately 3.5 mm in a cohort in which the patient age was controlled. The correlation was similar under mesopic conditions. The eye's tolerance to modulation transfer function decay was approximately 15% of the average modulation value.
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...