Precision of navigated stereotactic probe implantation into the brainstem

Henrik Giese1, Karl-Titus Hoffmann, Andreas Winkelmann

  • 1Department of Pediatric Neurosurgery, Charité Universitätsmedizin Berlin, Berlin, Germany.

Abstract

Insights

This study evaluated the VarioGuide frameless stereotactic system for brainstem probe placement. The system demonstrated good precision, supporting its use for biopsies and local chemotherapy in brainstem tumors.

Area of Science:

  • Neurosurgery
  • Medical Imaging
  • Biomedical Engineering

Background:

  • Debate exists regarding indications for stereotactic biopsies and local chemotherapy in diffuse brainstem tumors.
  • Precise probe placement is critical for the success of these brainstem interventions.

Purpose of the Study:

  • To evaluate the precision of brainstem probe positioning using a navigated frameless stereotactic system (VarioGuide).
  • To assess the accuracy of the VarioGuide system in experimental settings for potential clinical application in brainstem lesions.

Main Methods:

  • Utilized the VarioGuide stereotactic system to place probes in an agarose brain model and anatomical brainstem specimens.
  • Employed thin-sliced CT for planning, fused with volumetric T1-weighted MR imaging.
  • Measured deviations between planned and actual catheter trajectories post-intervention using CT and MR imaging.

Main Results:

  • In the agarose model, mean total target deviations were 2.8 +/- 1.2 mm (CT) and 3.1 +/- 1.2 mm (MR).
  • In anatomical specimens, suboccipital approach showed deviations of 1.95 +/- 0.6 mm (CT) and 1.8 +/- 0.7 mm (MR).
  • Precoronal approach in specimens yielded deviations of 2.2 +/- 1.2 mm (CT) and 2.1 +/- 1.1 mm (MR).

Conclusions:

  • The VarioGuide frameless stereotactic system demonstrates good precision for probe placement in deep-seated brainstem locations.
  • The system's accuracy supports its potential use for biopsies and probe implantation in patients with brainstem lesions.