Related Experiment Video
Updated: Jun 14, 2026

Applying X-ray Imaging Crystal Spectroscopy for Use as a High Temperature Plasma Diagnostic
Published on: August 25, 2016
Development of multiwavelength-range fine-resolution spectrometer for hydrogen emissions and its application to large
K Fujii1, K Mizushiri, T Nishioka
1Department of Mechanical Engineering and Science, Graduate School of Engineering, Kyoto University, Kyoto 606-8501, Japan.
Abstract:
We developed a spectrometer specialized for simultaneous observation of the hydrogen Balmer-alpha, -beta, -gamma lines and the Fulcher-alpha v(')=v(")=2 rovibronic transition band. The spectrometer was optimized for the light input coupled by nine optical fibers having 400 microm core diameters. The spectral resolutions were 0.02-0.03 nm for these wavelength ranges at the entrance slit width of 20 microm. The polarization resolved spectra of these emissions from the peripheral region of large helical device (LHD) plasmas were measured simultaneously and showed the polarization dependence coming from the magnetic field in the LHD plasma.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Emission Spectra
Atomic Emission Spectroscopy: Overview