Related Experiment Videos
Clinical and radiographic analysis of a partially cemented glenoid implant: five-year minimum follow-up
R Sean Churchill1, Christopher Zellmer, Herbert J Zimmers
1Aurora Advanced Healthcare, Milwaukee, WI 53209, USA. schurchill@ah.com
Hypothesis:
This study reviewed 20 consecutive shoulders that underwent total shoulder arthroplasty with a partially cemented glenoid component at a minimum follow-up of 5 years. We hypothesized that bone growth between the fins of the component would result in improved glenoid fixation and reduced lucent line formation.
Methods:
Patients were evaluated with the Simple Shoulder Test (SST), Constant scores, and radiographs. Postoperative radiographs taken at 6 weeks and 5 years were compared. Independent evaluators assigned every glenoid a lucency grade from 0 (no lucency) to 5 (gross lucency) and assessed the central peg for bone presence between the fins.
Results:
Mean SST and Constant scores were 11.1 and 82.4, respectively. Mean lucency scores increased from 0.1 to 0.3 from 6 weeks to the 5-year follow-up (P = .05). At the 5-year follow-up, 15 of 20 patients had similar or increased bone presence between the central peg fins, and none had a worsening lucency score. The 5 patients with decreased or absent bone presence had worsening lucency scores, from a mean of 0.2 to 1.0 (P = .05). There was no difference in SST score (P = .54) or Constant score (P = .37) between the two groups.
Conclusions:
This partially cemented glenoid component had low rates of lucency at the 5-year follow-up. Similar or increased bone growth between the fins of the central peg was associated with no progression of glenoid lucencies. Decreased or absent bone presence between the fins at follow-up correlated with slight advancement of glenoid lucent lines.