Dosimetry for Cell Irradiation using Orthovoltage (40-300 kV) X-Ray Facilities
Published on: February 20, 2021
Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film
J G H Sutherland1, D W O Rogers
1Ottawa Carleton Institute for Physics, Carleton University Campus, Ottawa K1S 5B6, Canada. jsutherl@physics.carleton.ca
Purpose:
The absorbed-dose energy dependence of GAFCHROMIC EBT and EBT2 film irradiated in photon beams is studied to understand the shape of the curves and the physics behind them.
Methods:
The absorbed-dose energy dependence is calculated using the EGSnrc-based EGS_chamber and DOSRZnrc codes by calculating the ratio of dose to water to dose to active film layers at photon energies ranging from 3 keV to 18 MeV. These data are compared to the mass energy absorption coefficient ratios and the restricted stopping power ratios of water to active film materials as well as to previous experimental results.
Results:
In the photon energy range of 100 keV to 18 MeV the absorbed-dose energy dependence is found to be energy independent within +/- 0.6%. However, below 100 keV, the absorbed-dose energy dependence of EBT varies by approximately 10% due to changes in mass energy absorption coefficient ratios of water to film materials, as well as an increase in the number of electrons being created and scattered in the central surface layer of the film. Results are found to disagree with previous experimental studies suggesting the possibility of an intrinsic energy dependence at lower photon energies. For EBT2 film the absorbed-dose energy dependence at low photon energies varies by 50% or 10% depending on the manufacturing lot due to changes in the ratio of mass energy absorption coefficients of the active emulsion layers to water.
Conclusions:
Caution is recommended when using GAFCHROMIC EBT/EBT2 films at photon energies below 100 keV. It is recommended that the effective atomic number of future films be produced as close to that of water and that thicker active layers are advantageous.
Related Concept Videos
Biological Effects of Radiation
Absorption of Radiation
Thomson's e/m Experiment
A particle with charge q, speed v, and mass m enters an area from the top, where the magnetic and electric fields are perpendicular both to the particle's motion and to one another. The magnetic...
Atomic Emission Spectroscopy: Overview
Adsorption Isotherms II
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...