Performance evaluation of diagnostic radiology dosimeters in clinical and calibration x-ray beams
Costantine John Hourdakis1, Argyro Boziari, Aggeliki Manetou
1Ionizing Radiation Calibration Laboratory, Greek Atomic Energy Commission, 15310 Agia Paraskevi, Athens, Attiki, Greece. khour@gaec.gr
Abstract:
Diagnostic radiology dosimeters should comply with International Electrotechnical Commission (IEC) 61674 standard in order to perform measurements with sufficient accuracy and reliability. The calibration of a dosimeter is performed under, and pertains to, reference conditions. However, in most cases, dosimeters are used for clinical measurements under non-reference conditions. The performance, in terms of accuracy of dose measurements, of six commercial diagnostic radiology dosimeters was tested at reference calibration and at clinical non-reference conditions. The results showed that all dosimeters being tested exhibited limits of variation within the +/-5% IEC limits. Depending on the detector's physical and operational properties, the dosimeters' energy dependence of response values varied from -4.7% to +4.2%. To address this variation of response, calibration at three radiation qualities (RQR 3, RQR 5, and RQR 9), at least, is recommended. Different irradiation conditions such as air kerma rate, x-ray tube design, x-ray system, and dosimeter operational modes affect the dosimeters' response by less than 3%. A dosimeter that complies with IEC standards and operates according to its specifications could be used at typical clinical irradiation conditions taking into account only corrections for the energy dependence of response. In this case, the error in dose accuracy is expected to be less than 3%.
Related Concept Videos
Radiological Investigation I: X-ray and CT
X-ray Imaging
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Biological Effects of Radiation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...