[Cause of micromotion in distal femoral prosthesis]
Wen-wei Dong1, Xue-zhong Yu, Yi-mu Guo
1The Second Affiliated Hospital, College of Medicine, Zhejiang University, Hangzhou 310009, China.
Objective:
To investigate possible causes of micromotion in distal femoral prosthesis.
Methods:
Based on the assumption that the femur and prosthesis were considered as concentric cylinders with completely bonded interface, a theoretical model simulating the interfacial stress transfer was established. The distributions of the interfacial shear and radial stresses with the changing of z were obtained through mathematics and mechanics deducing.
Results:
The maximum interfacial shear stress occurred at the position of z=0, namely, the cross section of the femur neck. The interfacial shear stress sharply decreased with the increasing of z and came to nearly zero at the range of z> 0.1 m. While the interfacial radial stress increased with the increasing of z,at the range of z >0.05 m it was constant and reached the maximum.
Conclusion:
The micromotion in distal prosthesis is caused by the interfacial radial stress.
More Related Videos
Related Concept Videos
Bones of the Lower Limb: Femur and Patella
Mechanism of Ciliary Motion
The cilia are made up of microtubules in a 9+2 arrangement, with nine microtubule doublet ring bundles, surrounding a pair of central singlet microtubule bundles. The doublet microtubule bundles are...
Muscles that Move the Leg
Anterior Compartment
The quadriceps femoris, the most visible muscle of the anterior compartment, is integral for leg extension and thigh flexion. It is formed by merging four distinct muscles — the vastus lateralis, vastus medialis, vastus intermedius, and rectus femoris. The quadriceps tendon, a shared tendon of the four quadriceps muscles, is affixed to...