Intercalibration of infrared channels of polar-orbiting IRAS/FY-3A with AIRS/Aqua data
1Key Laboratory of Wave Scattering and Remote Sensing Information (Ministry of Education), Fudan University, Shanghai 200433, PR China. jianggengming@hotmail.com
Abstract:
This work intercalibrated the infrared window channels 8 (12.47 microm), 9 (11.11 microm) and 19 (3.98 microm) of the InfraRed Atmospheric Sounder (IRAS) aboard the Chinese second generation polar-orbiting meteorological satellite FengYun 3A (FY-3A) with high spectral resolution data acquired by the Atmospheric InfraRed Sounder (AIRS) aboard Aqua. A North Pole study area was selected according to the IRAS and AIRS' viewing geometry. The IRAS/FY-3A L1 data and AIRS/Aqua 1B Infrared geolocated and calibrated radiances (AIRIBRAD) in July of 2008 were used in this work. A sub-pixel registration method was developed and applied to the IRAS and AIRS images to improve the intercalibration accuracy. The co-located measurement pairs were picked out with absolute Viewing Zenith Angle differences less than 5 degrees (|Delta VZA|<5 degrees), absolute Viewing Azimuth Angle differences less than 90 degrees (|Delta VAA|<90 degrees) and absolute time differences less than 15 min (|Delta T|<15'). The results reveal that the convolved AIRS/Aqua measurements are highly linearly related to the IRAS/FY-3A measurements with correlation coefficients greater than 0.93, and calibration discrepancies exist between IRAS and AIRS channels indeed. When the brightness temperatures in IRAS/FY-3A channels change from 230.0 K to 310.0 K, the AIRS-IRAS temperature adjustment linearly varies from -3.3 K to 1.7 K for IRAS/FY-3A channel 8, from -2.9 K to 2.6 K for IRAS/FY-3A channel 9, and from -5.3 K to 1.1 K for IRAS/FY-3A channel 19.
More Related Videos
Related Concept Videos
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...
IR Spectrometers
IR Frequency Region: X–H Stretching
IR Frequency Region: Fingerprint Region
IR Spectrum
Transmittance is defined as the ratio of the radiant power passing through a sample to that from the radiation's source. Multiplying the transmittance by 100 gives the percent transmittance (%T), which varies between 100% (no absorption) and 0%...