Related Experiment Video
Updated: Jun 13, 2026

Automated Two-dimensional Spatiotemporal Analysis of Mobile Single-molecule FRET Probes
Published on: November 23, 2021
Precision analysis for standard deviation measurements of immobile single fluorescent molecule images
Michael C DeSantis1, Shawn H DeCenzo, Je-Luen Li
1Department of Physics, Washington University, St. Louis, MO 63130, USA.
Abstract:
Standard deviation measurements of intensity profiles of stationary single fluorescent molecules are useful for studying axial localization, molecular orientation, and a fluorescence imaging system's spatial resolution. Here we report on the analysis of the precision of standard deviation measurements of intensity profiles of single fluorescent molecules imaged using an EMCCD camera.We have developed an analytical expression for the standard deviation measurement error of a single image which is a function of the total number of detected photons, the background photon noise, and the camera pixel size. The theoretical results agree well with the experimental, simulation, and numerical integration results. Using this expression, we show that single-molecule standard deviation measurements offer nanometer precision for a large range of experimental parameters.