Related Experiment Video
Updated: Jun 13, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
RP-HPLC Estimation of Venlafaxine Hydrochloride in Tablet Dosage Forms
S L Baldania1, K K Bhatt, R S Mehta
1Anand Pharmacy College, Opp. Town Hall, Anand - 388 001.
Abstract:
A simple, specific, accurate, and precise reverse phase high performance liquid chromatographic method was developed and validated for the estimation of venlafaxine hydrochloride in tablet dosage forms. A Phenomenex Gemini C-18, 5 mum column having 250 x 4.6 mm i.d. in isocratic mode, with mobile phase containing methanol: 0.05 M potassium dihydrogen orthophosphate (70:30, v/v; pH 6.2) was used. The flow rate was 1.0 ml/min and effluents were monitored at 226 nm. Carbamazepine was used as an internal standard. The retention time of venlafaxine hydrochloride and carbamazepine were 3.7 min and 5.3 min, respectively. The method was validated for specificity, linearity, accuracy, precision, limit of quantification, limit of detection, robustness and solution stability. Limit of detection and limit of quantification for estimation of venlafaxine hydrochloride were found to be 100 ng/ml and 300 ng/ml, respectively. Recoveries of venlafaxine hydrochloride in tablet formulations were found to be in the range of 99.02-101.68%. Proposed method was successfully applied for the quantitative determination of venlafaxine hydrochloride in tablet dosage forms.